Accéder au contenu
Merck

GC Analysis of BTEX, Undecane, and Tridecane on SLB®-IL100

GC Analysis of BTEX, Undecane, and Tridecane on SLB®-IL100 suitable for GC

Matériaux

Colonne analytique

Réf. du produit
Description
Tarif

SLB®-IL100 Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.20 μm

CONDITIONS

column

SLB-IL100, 30 m × 0.25 mm I.D., 0.20 μm (28884-U)

oven

110 °C

inj. temp.

250 °C

detector

FID, 250 °C

carrier gas

helium, 26 cm/sec

injection

0.1 μL, 300:1 split

liner

4 mm I.D., split, cup design

sample

neat mixture containing varying percentages of each component

Description

Informations légales

SLB is a registered trademark of Merck KGaA, Darmstadt, Germany