Saltar al contenido
Merck

Identification of diethylene glycol monobutyl ether as a source of contamination in an ion trap mass spectrometer.

Journal of the American Society for Mass Spectrometry (2003-11-05)
Christopher R Gibson, Cheri M Brown
RESUMEN

Tandem liquid chromatography-mass spectrometry coupled to online radioactive material detection (LC/RAM/MS/MS) is a technique that is used routinely for in vivo and in vitro drug metabolism studies and allows for a simultaneous correlation between radiochemical peaks and mass spectral data. The compound diethylene glycol monobutyl ether (DGBE), a component of a commercially available scintillation cocktail for RAM analysis, was identified as a source of overwhelming chemical noise in a mass spectrometer which was used in an LC/RAM/MS/MS configuration. In this report, we describe the identification of DGBE as the source of the chemical noise and the methods that were used to minimize the exposure of the mass spectrometer to volatile components of the scintillation cocktail.

MATERIALES
Referencia del producto
Marca
Descripción del producto

Sigma-Aldrich
Diethylene glycol butyl ether, ≥99%
Sigma-Aldrich
Diethylene glycol monobutyl ether, ≥98.0% (GC)
Supelco
Diethylene glycol monobutyl ether, for surfactant analysis, ≥99.0%
Sigma-Aldrich
Diethylene glycol butyl ether, SAJ special grade, ≥99.0%