HPLC Analysis of Phenolics from Rosemary on Ascentis® Express C8
Matériaux
Colonne analytique
Réf. du produit
Description
Tarif
CONDITIONS
column
Ascentis Express C8, 5 cm x 3 mm I.D., 2.7 μm particles (53848-U)
mobile phase
[A] 0.1% formic acid in water; [B] 0.1% formic acid in acetonitrile
gradient
15 to 100% B in 8.5 min
flow rate
0.6 mL/min
column temp.
35 °C
detector
UV, 214 nm
injection
2 μL
sample
100 mg/L in 75 μM EDTA, 1.5 mM ascorbate, 15% acetonitrile
Description
Remarque sur l'analyse
This application demonstrates the suitability of Ascentis Express C8 for the efficient separation of phenolics from rosemary.
Informations légales