Direkt zum Inhalt
Merck

HPLC Analysis of Chinese Ginseng Extract on Ascentis® Express C18

HPLC Analysis of Chinese Ginseng Extract on Ascentis® Express C18 application for HPLC

Materialien

Analytische Säule

Produkt-Nr.
Beschreibung
Preisangaben

Ascentis® Express C18, 2,7 μm HPLC-Säule

2.7 μm particle size, L × I.D. 15 cm × 4.6 mm

CONDITIONS

column

Ascentis Express C18, 15 cm x 4.6 mm I.D., 2.7 μm particles (53829-U)

mobile phase

[A] water; [B] acetonitrile

gradient

held at 25% B for 1.5 min; 25 to 85% B in 12 min; held at 85% for 1 min

flow rate

1 mL/min

pressure

3481 psi (240 bar)

column temp.

60 °C

detector

UV, 205 nm

injection

25 μL

sample

Extracted from dried, ground plant material; 10 mg/mL in water:acetonitrile (85:15); sonicate 10 min at ambient

Beschreibung

Hinweis zur Analyse

This application demonstrates the suitability of Ascentis Express C18 for the efficient separation of Chinese ginseng root extract.

Rechtliche Hinweise

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany