GC Analysis of PCBs in Whole Cow Milk (Spiked) on SLB®-5ms (20 m x 0.18 mm I.D., 0.18 μm) after QuEChERS Cleanup using Supel QuE Z-Sep, Fast GC Analysis

GC Analysis of PCBs in Whole Cow Milk (Spiked) on SLB®-5ms (20 m x 0.18 mm I.D., 0.18 μm) after QuEChERS Cleanup using Supel™ QuE Z-Sep, Fast GC Analysis suitable for GC, application for SPE

Materiales

Analytical column

Referencia del producto
Descripción
Precios

SLB®-5ms Capillary GC Column

L × I.D. 20 m × 0.18 mm, df 0.18 μm

SPE tube or plate

Referencia del producto
Descripción
Precios

Supel QuE

Empty Centrifuge Tube with Lid, centrifuge tube volume 50 mL, pk of 50, suitable for EN 15662:2008 per BS

Supel QuE

Non-Buffered Tube 1, pk of 50

Supel QuE

Z-Sep+ Bulk, pkg of 20 g

standard

Referencia del producto
Descripción
Precios

PCB standard solution 7

analytical standard, 10 μg/mL in isooctane

CONDITIONS

sample preparation

SPE (Solid Phase Extraction)

sample/matrix

10 g whole cow′s milk

extraction tube

Supel QuE Non-buffered tube (4 g magnesium sulfate, 1 g sodium chloride) (55294-U)

clean-up tube

Supel QuE Z-Sep 150 mg, magnesium sulfate (150 mg, part no. 63135) (55299-U)

column

SLB-5ms, 20 m x 0.18 mm I.D., 0.18 μm (28564-U)

oven

75 °C (1 min), 12 °C/min to 340 °C (10 min)

inj. temp.

250 °C

detector

μ-ECD, 340 °C

carrier gas

hydrogen, 1.2 mL/min, constant

injection

1 μL splitless (0.75 min)

liner

4 mm I.D. FocusLiner with taper

Descripción

General description

Reference: Stenerson, K.; Ye, M.; Halpenny, M., New Analytical Tools for the Determination of Persistent Organic Pollutants (POPs) in Fatty Food and Beverage Matrices Using QuEChERS Extraction/Cleanup and Gas Chromatography (GC) Analysis. Pittcon 2013. Sigma-Aldrich/Supelco publication T413034

Analysis Note

This application demonstrates the use of dispersive SPE or "QuEChERS" for the extraction and cleanup of whole cow′s milk for GC-ECD analysis of polychlorinated biphenyl (PCBs). Z-Sep sorbent was used in the cleanup step and found to reduce background and yield recoveries >70% for all PCB congeners evaluated except no. 209, which was recovered at 68%.
Extraction solvent: acetone:hexane (1:1), 20 mL
Clean-up volume: 1 mL