Accéder au contenu
Merck

HPA dans l′eau

HPA dans l’eau suitable for GC, application for SPE

Matériaux

Tube ou plaque pour SPE

Réf. du produit
Description
Tarif

Tube SPE Supelclean ENVI-18

bed wt. 500 mg, volume 6 mL, pkg of 30 ea

CONDITIONS

sample preparation

SPE (Solid Phase Extraction)

sample/matrix

water spiked with PAHs

SPE tube/cartridge

Supelclean ENVI-18, 500 mg/6 mL (57064)

condition

2 x 6 mL toluene:methanol (10:1); 6 mL methanol; 6 mL DI water

sample addition

250 mL water sample at 10 mL/min

drying

10 min under vacuum

elution

2 x 1 mL toluene:methanol (10:1)

column

5% phenyl polysiloxane, 30m × 0.25 mm I.D., 0.25 μm film (Supelco equivalent, SLB-5ms, 28471-U)

oven

70 °C (2 min) to 280 °C at 8 °C/min

detector

FID, 310 °C

carrier gas

helium

injection

1 μL

Description

Informations légales