Accéder au contenu
Merck

HPLC Analysis of Steroids on Ascentis® Express C18 (10 cm x 2.1 mm I.D.)

HPLC Analysis of Steroids on Ascentis® Express C18 (10 cm x 2.1 mm I.D.) application for HPLC

Matériaux

Colonne analytique

Réf. du produit
Description
Tarif

Colonne de CLHP Ascentis® Express C18, 2,7 μm

2.7 μm particle size, L × I.D. 10 cm × 2.1 mm

Étalon

Réf. du produit
Description
Tarif

Testosterone

≥98%

Epitestosterone solution

1.0 mg/mL in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®

Methandienone solution

1.0 mg/mL in 1,2-dimethoxyethane, ampule of 1 mL, certified reference material, Cerilliant®

Nandrolone solution

1.0 mg/mL in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®

Norethandrolone solution

1.0 mg/mL in 1,2-dimethoxyethane, ampule of 1 mL, certified reference material, Cerilliant®

CONDITIONS

column

Ascentis Express C18, 10 cm x 2.1 mm I.D., 2.7 μm particles (53823-U)

mobile phase

[A] water; [B] acetonitrile; (60:40, A:B)

flow rate

0.2 mL/min

pressure

2944 psi (203 bar)

column temp.

35 °C

detector

UV, 254 nm

injection

2 μL

sample

50 mg/L in 75:25, water:methanol

Description

Remarque sur l'analyse

This application demonstrates the suitability of Ascentis Express C18 for the efficient separation of 5 steroids.

Informations légales

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany