796-0315
US EPA Method 0030: GC Analysis of Volatiles on SPB®-624 after Collection/Desorption using VOST 100 Tube
suitable for GC, application for air monitoring
About This Item
technique(s)
gas chromatography (GC): suitable
test parameters
sample: calibration standard: volatiles in methanol; blank: Tenax TA tube with surrogate thermally desorbed into purge & trap system
adsorbent tube: VOST 100 (20074-U)
purge trap: VOCARB 4000 (20308)
column: SPB-624, 75 m × 0.53 mm I.D., 3.0 μm film (25432)
oven: 35 °C (5 min) to 200 °C at 10 °C/min, hold 2 min
detector: FID, 300 °C
carrier gas: helium, 8 mL/min
suitability
application for air monitoring
application(s)
air monitoring
environmental
industrial hygiene