Accéder au contenu
Merck

US EPA Method TO-17: GC Analysis of Volatiles on VOCOL after Collection/Desorption using Carbotrap® 300 Tube

US EPA Method TO-17: GC Analysis of Volatiles on VOCOL after Collection/Desorption using Carbotrap® 300 Tube suitable for GC, application for air monitoring

Matériaux

Colonne analytique

Réf. du produit
Description
Tarif

SPB-1®

L × I.D. 60 m × 0.25 mm, df 3.00 μm

Produit(s) apparenté(s)

Réf. du produit
Description
Tarif

Carbotrap® 300

glass TD tube, fritted, O.D. × I.D. × L 6 mm × 4 mm × 7 in., preconditioned, pkg of 1 ea

Étalon

Réf. du produit
Description
Tarif

JHAP-43 Mix

1 ppm in nitrogen, pkg of 110 L (pi-marked cylinder), analytical standard

CONDITIONS

sample/matrix

JHAP-43 Gas Mix (500429), 1ppm each component, 20mL

adsorbent tube

Carbotrap 300 (28283-U)

desorption process

20 mL/min at 330 °C for 6 min (cryo-focused to inlet @ -150°C)

column

SPB-1, 60 m × 0.25 mm I.D., 3.0 μm

oven

35°C (8 min), 5°C/min to 100°C (0 min), 15°C/min to 230°C (8 min)

detector

MSD, scan range 35-269

carrier gas

helium 31cm/sec

Description

Informations légales

Carbotrap is a registered trademark of Merck KGaA, Darmstadt, Germany