Skip to Content
MilliporeSigma

LC/MS Analysis of Cetirizine and Hydroxyzine in Urine on Ascentis® Express F5

LC/MS Analysis of Cetirizine and Hydroxyzine in Urine on Ascentis® Express F5 application for LC-MS

Materials

analytical column

Product No.
Description
Pricing

Ascentis® Express F5, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 10 cm × 3 mm

standard

Product No.
Description
Pricing

Cetirizine dihydrochloride solution

1.0 mg/mL in acetonitrile: water (9:1) (as free base), ampule of 1 mL, certified reference material, Cerilliant®

Hydroxyzine dihydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

CONDITIONS

column

Ascentis Express F5, 10 cm x 3.0 mm I.D., 2.7 μm particles (53578-U)

mobile phase

5 mM ammonium formate pH 3.0 with formic acid in 90:10, acetonitrile:water

flow rate

0.6 mL/min

pressure

1455 psi (100 bar)

column temp.

25 °C

detector

MS, ESI+, combined SIR

injection

1 μL

sample

100 ng/mL in urine dilute 1:1 with water

Description

Legal Information

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany