US EPA Method 8015 (modified): GC Analysis of Gasoline Range Organics (GRO) on VOCOL® after Purge & Trap using "M" Trap

suitable for GC

Sign Into View Organizational & Contract Pricing


About This Item

technique(s)

gas chromatography (GC): suitable

test parameters

sample/matrix: Each analyte at 50 ppb in 5 mL water
purge trap: M trap (20079-U)
purge: 40 mL/min at 25 °C for 11 min
dry purge: 2 min
desorption preheat: 220 °C
desorption process: 40 mL/min at 225 °C for 2 min
bake: 260 °C for 10 min
transfer line/valve temp.: 110 °C
column: VOCOL, 30 m x 0.25 mm I.D., 1.5 μm (24205-U)
oven: 60 °C (2 min), 8 °C/min to 165 °C, 25 °C/min to 220 °C (5 min)
detector: FID, 250 °C
carrier gas: helium, 1.0 mL/min
injection: 30:1 split
sample: 0.75 mm I.D. SPME
inj. temp.: 200 °C

application(s)

environmental

Legal Information

VOCOL is a registered trademark of Merck KGaA, Darmstadt, Germany

Analyte

Description

    Pentane
    2-Methylpentane
    2,2,4-Trimethylpentane
    Benzene
    Toluene
    Nonane
    Ethylbenzene
    m-Xylene
    p-Xylene
    o-Xylene
    Decane
    1,2,4-Trimethylbenzene
    Butylcyclohexane
    Naphthalene
    See All (14)

Choose from one of the most recent versions:

Certificates of Analysis (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Customer Support.

Already Own This Product?

Find documentation for the products that you have recently purchased in the Document Library.

Visit the Document Library

Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.

Contact Technical Service