Pular para o conteúdo
Merck
Todas as fotos(1)

Documentos

APP501

HPLC Analysis of Cinacalcet Hydrochloride Impurities with a Purospher® STAR Phenyl Column

Faça loginpara ver os preços organizacionais e de contrato


About This Item

técnica(s)

HPLC: suitable

parâmetros de teste

column: Purospher® STAR Phenyl (5 µm) LiChroCART® 150 x 4.6 mm I.D. (1.51922)
mobile phase: [A] 10 mmol/L Sodium perchlorate solution, adjusted to pH 2.5 with perchloric acid; [B] Acetonitrile
gradient: 0-0.5 min 2.5% B, in 4.5 min to 95%B, 2 min 95%B, 3 min at 2.5%B
flow rate: 1.0 mL/min
pressure: 60 bar (870 psi)
column temp.: 60°C (Sampler: 10 °C)
detector: UV, 215 nm (analytical flow cell; 10 µL)
injection: 5 µL
sample/matrix: Impurity identification solution with cinacalcet impurities A, B, C at 1.5 µg/mL each and cinacalcet at 5 µg/mL in solvnt mixture (10 mmol/L sodium perchlorate solution, adjusted to pH 2.5 with perchloric acid, and acetonitrile in a ratio of 50:50 (%v/v))

aplicação(ões)

pharmaceutical

Descrição geral

Cinacalcet hydrochloride is an oral calcimimetic indicated for the treatment of secondary hyperthyroidism in patients on dialysis with end-stage renal disease, and in patients with parathyroid carcinoma to reduce hypercalcemia.

Nota de análise

The developed method on a Purospher® STAR Phenyl, 150 × 4.6 mm (5µm) column was found to be linear for cinacalcet hydrochloride up to a concentration of 6 µg/mL with a LOQ of 0.65 µg/mL and LOD of 0.21 µg/mL. For cinacalcet impurity A, impurity B and impurity C, the method was linear up to a concentration of 1.8 µg/mL with LOQ ranging from 0.03 to 0.13 µg/mL and LOD ranging from 0.01 to 0.04 µg/mL. The method can be applied for the study of related substances in cinacalcet hydrochloride API using the Purospher® STAR Phenyl HPLC column. Repeatability for the cinacalcet 5 µg/mL solution was 1.19% RSD (n=5). The tailing factor for cinacalcet eluting at 12.3 min was NMT 2.0 (1.31) and the theoretical plate count was significantly above 2000.

Outras notas

App_501

Informações legais

LICHROCART is a registered trademark of Merck KGaA, Darmstadt, Germany
Purospher is a registered trademark of Merck KGaA, Darmstadt, Germany

Escolha uma das versões mais recentes:

Certificados de análise (COA)

Lot/Batch Number

Lamentamos, não temos COA para este produto disponíveis online no momento.

Se precisar de ajuda, entre em contato Atendimento ao cliente

Já possui este produto?

Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.

Visite a Biblioteca de Documentos

Development and validation of a stability-indicating RP-UPLC method for the estimation of impurities in cinacalcet hydrochloride API and its formulation
Reddy PS, et al.
Scientia Pharmaceutica, 83(4) (2015)

Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.

Entre em contato com a assistência técnica