ZAFS0080WW
AFS® Water Purification System
For analyzers with pure water needs up to 160 L/day
Sinônimo(s):
AFS® 8 Water Purification System
About This Item
Produtos recomendados
entrada CA/CC
100 - 230 V, 50 - 60 Hz
Nível de qualidade
linha de produto
AFS®
embalagem
pkg of 1 unit
Parâmetros
160 L/day max. usage
2 L/min distribution flow rate
28 kg operating weight (61.7 lb)
8 L/hr make-up flow rate
resistividade
>15 MΩ-cm, at 25 degrees
system H × W × depth
58.5 cm (23 in.) × 26.8 cm (10.6 in.) × 42.6 cm (16.8 in.)
resíduo de carbono orgânico total (TOC)
<30 ppb
entrada
potable tap water
saída
clinical laboratory reagent water (CLRW) (complies with the CLSI guideline)
Categorias relacionadas
Descrição geral
Aplicação
Características e benefícios
- Complementary water purification techniques provide consistent water quality meeting CLRW standards
- Cost-effective thanks to optimized lifetimes for pretreatment and polishing packs and low feed water use
- Fulfills accreditation needs (e.g. CAP 15189SM accreditation to the ISO 15189:2022 standard) with full monitoring and automatic water quality archiving capabilities. Up to 6 months of information can be stored for reliable traceability
- An intuitive icon-based system display facilitates maintenance, and an ergonomic pack-locking system makes pack changes easier than ever
- Milli-Q® Service Plans offer a range of support, including preventive maintenance visits and online contract management
- A small footprint allows installation wherever it’s convenient on the wall, on or under the bench, or on a cart
Componentes
- The indicated price currently corresponds to the above catalog number only; it is subject to change at any time without notice.
- This catalog number does not include consumables, accessories and other products and/or services that may be required for the installation and the correct operation of the water purification unit over time.
- Because our customers each have specific purified water needs, we recommend that you have your purchase validated by one of our specialists. We remain at your disposal to advise you, do not hesitate to contact us!
Outras notas
Organism Retention: Microorganisms
Mode of Action: Filtering and UV
Application: Clinical laboratories
Intended Use: Water purification
Instructions for Use: This item provides water filtered through 0.22 μm sterilizing-grade membranes and exposed to bactericidal UV lamp. Please refer to system equipment user guide section "System" (provided on USB key in the North American configuration version).
Storage Statement: Store in dry location
Disposal Statement: Dispose of in accordance with applicable federal, state, and local regulations.
Informações legais
Certificados de análise (COA)
Busque Certificados de análise (COA) digitando o Número do Lote do produto. Os números de lote e remessa podem ser encontrados no rótulo de um produto após a palavra “Lot” ou “Batch”.
Já possui este produto?
Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.
Artigos
Critical points for laboratories to consider when assessing bacteria levels in purified water and attempting to optimize the microbiology process.
Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.
Entre em contato com a assistência técnica