Accéder au contenu
Merck

LC/MS Analysis of Omeprazole and Metabolites on Ascentis® Express RP Amide (pH 2.7)

LC/MS Analysis of Omeprazole and Metabolites on Ascentis® Express RP Amide (pH 2.7) application for HPLC

Matériaux

Colonne analytique

Réf. du produit
Description
Tarif

Ascentis® Express RP-Amide, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 5 cm × 2.1 mm

Composant de phase mobile

Réf. du produit
Description
Tarif

Formate d′ammonium

eluent additive for LC-MS, LiChropur, ≥99.0%

Étalon

Réf. du produit
Description
Tarif

Phenacetin

≥98.0% (HPLC)

Omeprazole sulfone

analytical standard

Omeprazole

analytical standard

CONDITIONS

column

Ascentis Express RP-Amide, 5 cm x 2.1 mm I.D., 2.7 μm particles (53911-U)

mobile phase

[A] 10 mM ammonium formate in water, pH 2.7; [B] 10 mM ammonium formate in methanol, pH 2.7; (65:35, A:B)

flow rate

0.3 mL/min

column temp.

35 °C

detector

ESI(+), m/z 50-400

injection

1 μL

sample

1 mg/L in mobile phase

Description

Remarque sur l'analyse

This application demonstrates the suitability of the Ascentis Express RP Amide for the analysis of omeprazole metabolites.

Informations légales