HPLC Analysis of Phenyl Urea Pesticides on Ascentis® Express C18 Phase
Matériaux
Colonne analytique
Réf. du produit
Description
Tarif
CONDITIONS
column
Ascentis® Express C18, 10 cm x 4.6 mm I.D., 2.7 μm particles (53827-U)
mobile phase
[A] 0.025 M potassium phosphate buffer, adj. to pH = 2.5, [B] acetonitrile; (50:50, A:B)
flow rate
2.0 mL/min
pressure
300 bar
column temp.
30 °C
detector
UV 245 nm, VWD
injection
0.5 μL
sample
acetonitrile
Description
Description générale
This separation illustrates the use of the highly efficient 2.7 um Ascentis Express Fused-Core C18 stationary phase in the analysis of some popular herbicides. The short run times allow analyses using isocratic conditions so that column equilibration time is not required between runs.
Autres remarques
Response Time: 0.02 sec.