Skip to Content
Merck
  • An Optical Fiber Chemical Sensor for the Detection of Copper(II) in Drinking Water.

An Optical Fiber Chemical Sensor for the Detection of Copper(II) in Drinking Water.

Sensors (Basel, Switzerland) (2019-12-05)
Maria Pesavento, Antonella Profumo, Daniele Merli, Lucia Cucca, Luigi Zeni, Nunzio Cennamo
ABSTRACT

Highly sensitive plasmonic optical fiber platforms combined with receptors have been recently used to obtain selective sensors. A low-cost configuration can be obtained exploiting a D-shaped plastic optical fiber covered with a multilayer sensing surface. The multilayer consists of a gold film, functionalized with a specific receptor, where the surface plasmon resonance (SPR) occurs. The signal is produced by the refractive index variation occurring as a consequence of the receptor-to analyte binding. In this work, a selective sensor for copper(II) detection in drinking water, exploiting a self-assembled monolayer (SAM) of d,l-penicillamine as the sensing layer, has been developed and tested. Different concentrations of copper(II) in NaCl 0.1 M solutions at different pH values and in a real matrix (drinking water) have been considered. The results show that the sensor is able to sense copper(II) at concentrations ranging from 4 × 10-6 M to 2 × 10-4 M. The use of this optical chemical sensor is a very attractive perspective for fast, in situ and low-cost detection of Cu(II) in drinking water for human health concerns. Furthermore, the possibility of remote control is feasible as well, because optical fibers are employed.

MATERIALS
Product Number
Brand
Product Description

Sigma-Aldrich
L-Penicillamine, 99%