Skip to Content
Merck

US EPA Method 610: GC Analysis of Polynuclear Aromatic Hydrocarbons (PAHs) on SPB®-608

US EPA Method 610: GC Analysis of Polynuclear Aromatic Hydrocarbons (PAHs) on SPB®-608 suitable for GC

Materials

analytical column

Product No.
Description
Pricing

SPB®-608 Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.25 μm

CONDITIONS

column

SPB-608, 30 m x 0.25 mm I.D., 0.25 μm (24103-U)

oven

70 °C (0.2 min), 25 °C/min to 265 °C, 5 °C/min to 300 °C (12 min)

inj. temp.

250 °C

detector

FlD, 300 °C

carrier gas

helium, 1.3 mL/min, constant

injection

2.0 μL pulsed splitless (30 psi until 0.2 min)

liner

4 mm I.D. single taper

sample

PAH standard, 1 μg/mL in methylene chloride

Description

Analysis Note

US EPA Method 610 describes the analysis of polynuclear aromatic hydrocarbons (commonIy referred to as PAHs or PNAs) by both HPLC and GC. The method designates the use of a packed GC column, however capillary columns will provide better resolution, baseline stability and longer lifetime. This application illustrates the separation of the target compound list in this method on the SPB-608 capillary GC column. Overall analysis time on the SPB-608 is longer than the SLB-5ms due to stronger retention and its lower maximum operating temperature. A pressure pulsed injection was used to improve response of the heavier PAHs, and to reduce the larger vapor volume resulting from a 2 μL injection.

Legal Information

SLB is a registered trademark of Merck KGaA, Darmstadt, Germany